The Transient Waveform Recorder (TWR) is designed to improve the
energy resolution for high energy phenomena such as Extreme High Energy
Neutrino, High Energy Cascade and Gama Ray Burst.
During the last pole summer season (Dec 2001 - Feb 2002) the prototypes
of 48 channels are installed (green in the figure). All remaining
channels
will be installed in this summer season (Dec 2002 - Feb 2003). In
this
note we will describe an analysis for charge measurement using the TWR
data. More information about TWR system can be found in TWR home
page.
For this analysis we use the N2 Laser data that have been taken for the calibration purpose. The source of N2 laser is located close to OM 35 on string 5 (blinking in the figure). The intensity of laser light can be adjusted by changing the attenuation factor. The following summarizes the information of the data set used and detailed information can be found in Andrea's note.
|
Run Number |
Attenuation |
|
R5332 |
ND 3.0 |
|
R5333 |
ND 3.0 |
|
R5334 |
ND 2.0 |
|
R5335 |
ND 1.0 |
|
R5336 |
ND 0.5 |
|
R5337 |
ND 0.0 |
The data recorded by the TWR DAQ is merged into the standard AMANDA DAQ
data by matching the event times of these two DAQ systems (wavemerger).
To eliminate the atmosphere muon events from N2 laser data we require
that event should have reasonable number of waveforms, which is similar
to the
Nch variable in normal data analysis. Figure 1 shows the distribution
of
number of waveform (Nwf) for R5337. The bump in the lower Nwf
region is corresponding to non-N2 laser event like the atmosphear muon
event.
So we require Nwf > 25.
Figure 1 : Number of waveform distribution.
Figure 2 shows an example of waveform for OM 360. Due to the long distance between laser source and OM only few number of photons can reach the PMT. There are clear two signal pulses in the figure. Each pulse corresponds to the single photoelecton signals. In the current TWR system one channel records 1024 bins for an event, where one bin is corresponding to 10 ns time. The figure 2 is just a part of the full waveform after removing no signal parts. Small size of overshoots are observed in the tail parts of pulses.
Figure 3 shows another example waveform of OM 662 which is located in
more bright position (distance from laser source is d=176 m).
Here we can still
distinguish
individual pulses for each photoelectons. Larger size of overshoot than
OM 360 is seen. Each signals are close so that the baseline of later
signal
can be shifted by the overshoot of earlier pulse.
Figure 4 shows very bright example of OM 621 (d=145m). Due to the short
flight distance of the incoming photons, there is large number of
photoelectrons signals with the narrow time spread. Now it is almost
impossible to count each photoelectron.
Here the baseline shift
caused
by overshoot is significantly larger than OM 662. An after pulse is
also
observed around 4 micro second after the primary pulse.
An extremely bright example is shown in Figure 5. This is the
waveform of OM 500 which is located at a distance d=54 m from the light
source.
Figure 5 : Waveform of OM 500.
Even though we expect huge and clean signal in this
bright OM, only a part of the signal is shown in this wave form. The
waveform
is saturated by the maximum voltage limitation of the ORB prompt
output.
A very broad pulse appears about after 2 micro second after the primary
peak.
This peak is not contributed by real photoelectron, but is caused by
the
discharge procedure of the capacitor. Also here the after pulses are
clearly
shown at about 4-5 micro second after the primary pulse.
Figure 6 : Fixed baseline and moving baseline method
One of the simple way to measure the charge from the waveform is the fixed threshold method. The charge is easily obtained by integrating the waveform under the fixed threshold. This method is similar to that for obtaining the peak ADC in the muon DAQ. We can have the stable and quick software using only a few lines of code. However this method has the problem of a wrong charge reconstruction due to the baseline shift caused by the overshoot and the fake peak of the discharging process. The moving baseline method can solve this problem. If we can trace the correct baseline under the signal pulses, the moving threshold can be obtained by taking an offset from this baseline. Then charge is calculated by integration the waveform over the threshold. However this is more complicated than the fixed threshold method.
The most important feature of the baseline method is how to find the
correct baseline. The basic idea of the baseline finder is based
on a cleaning procedure. We remove all peaked data points in the
waveform, then the baseline is obtained by connecting the remaining
points. Figure 7 briefly explains how to get the baseline.
First, we remove the points with large deviation. The
deviation of the given point is determined by the voltage difference
from its neighboring points. If these deviations are larger than 30 mV
the point is removed.
Basically most of peaked points are well cleaned by this procedure (See
Figure 7b).
Second, we remove the "clusters" with large deviation.
As it can be seen from Figure 7b, there are still remaining points
after
the first cleaning procedure, which are usually appeared in the plateau
region
of signal peaks. To remove these plateau points, we construct the
clusters
by connecting the remaining continuous data points. Then we remove the
cluster if the average voltage of cluster is largely deviated from
those
neighboring clusters.
Third, we remove the tail part of signal peak. The
remaining
data points in the tail part of the signal causes the unstable
fluctuation
of the baseline. More smooth baseline can be obtained by removing this
tail
components (See Figure 7c). Because of the long falling time of
PMT
signal, the data points in this part usually do not have large voltage
deviations
so that the previous cuts are not so effective to remove it. We
developed
a simple pattern recognition technique to find tail structure. We
1) take the four continuous points; 2) compare those deviations
(slopes);
3) if the first deviation is larger than 10mV and the slopes are
decreasing
as the time increase, the four points are removed from the baseline
calculation (Figure 7d).
Fourth, the baseline is obtained by connecting all
remaining points.
In order to reduce the statistical fluctuation, we calculate the
average
voltage for each 80 ns bins and then make a average line connecting
those
bins.
Finally, charge is simply calculated by integrating the waveform
under
the baseline.
Figure 7a : Example waveform which has two signal peaks.
Figure 7b : Waveform after removing the points with large deviation.
A few points in peak can be remained (in the red circle)
Figure 7c : Waveform after removing the cluster with large deviation
The tail components of each signal peaks are still remained.
Figure 7d : Waveform after cleaning tail component.
Now base line can be obtained by connecting all remaining points
Figure 7e : Waveform and obtained baseline.
Charge is calculated by integrating the waveform under the baseline.
As some examples are shown in Figure 8 the baseline finder looks well running. Baselines are reasonably obtained not only for the simple waveforms which contain one or two signal peaks, but also for complicated waveforms consisted number of photoelectrons . The baseline shift due to the overshoot, and even oscillated baseline, are also well corrected. The fake signal from the intermediate discharge procedure is well traced by our program.
Figure 8 : Waveforms are obtained baseline
Time scales are different.
Where blue line and circle is original waveform, red line is the
obtained baseline
and color filled circles indicate signal peak over the threshold (20mV)
;
green circle is leading edge, yellow is trailing edge and red is for
others.
More example are available :
309, 318, 323, 329, 333, 350, 360, 371, 375, 397,
402, 405, 431, 441, 445, 450, 458, 464, 473, 478,
487, 500, 506, 511, 514, 525, 529, 537, 542, 552,
597, 605, 611, 621, 625, 630, 634, 641, 657, 662,
665, 667, 668, 669, 671, 672, 673, 674,
To verify our method, we test the program with Monte Carlo. MC laser
samples are generated by the AMASIM with the PHOTONICS. The number of
photons is found to be 1.3*10^11 for the ND=0.0 laser source. This
value is determined by comparing the Nwf distribution
between MC and Data.
Figure 9 shows the measured Q distribution as a function of the arrival
number of photoelectron (Npe) which contribute to make signal pulse in
PMT. Where Q is the integrated charged for given waveform which
is
obtained by the moving baseline method. And Npe is by MC generator
information.
A Good linearity between the Npe and the measured Q is shown
(around
the red line in the figure) upto about 2000 Npe. The OMs with
different
slopes in the figure are due to the different gains of OMs. Also
two
OMs with huge Npe show clear saturation.
Figure 9 : Measured Q distribution vs. as Npe.
The Qs are measured by the moving threshold method.
Figure 10 shows the peak ADC distribution as a function of Npe, here the peak ADC is the pulse height recorded in HIT information (with muon DAQ). The saturations are started from OM with Npe=50~100. It indicates that TWR system can improve the dynamic range by more than factor few tens.
Figure 10 : Peak ADC vs. Npe.
Scale of x-axis is just 1/10 of Figure 9
Figure 11 compares two methods of the fixed threshold method and the
moving threshold method. Except some OMs having large overshoot
characteristic, basically there are good linearity between results of
two methods. Because the MC baselines are much stable than real data,
two methods should give us similar results in MC.
Figure 11 : Q Moving Threshold vs. Q Fixed Threshold.
Figure 12 shows an example of measured Q distributions obtained in the N2 laser Data. Where the laser intensity is ND=0.0 and OM number is 662. For the peak ADC distribution from Mu DAQ system, the saturation is clearly visible at about 2000 mV. Instead of this there are no saturation in the TWR measurements. Furthermore the distributions of TWR measurement are reasonably symmetric. Since the overshoot of this OM is not so serious, the results from both methods are comparable. The moving threshold method reduces the tail of the lower Q region in the fixed threshold method. While it makes the a little longer tail the upper Q side. The relative spreads (rms/mean) of two methods are almost same.
Figure 12 : Measured Q distributions of OM 662 with N2 laser
Data.
How TWR system increases the dynamic range is well shown in Figure 13. It is a scatter plot of the measured Q with TWR vs. peak ADC with muon DAQ. The moving threshold method is used for TWR measurement. The good linearity in lower Q region is broken at ADC~2000mV or 3500mV. These saturation voltages are channel dependent. However we should note it does not mean that the Q using TWR is correct.
Figure 13 : Q from TWR(moving threshold) vs. Peak ADC from Mu DAQ
In the real data it's difficult to verify our method since the true Npe
is not available. However we still have some indirect method.
Since the optical property of south pole ice is well measured, we can
expect the number of arrival photons using the attenuation relation,
Nph=N0 1/r e-r/lambda,
where, Nph is number of arrived photons, N0 is number of emitted photons at laser source, r is the distance between laser source and OM, and lambda is effective attenuation length. Assuming the Q is proportional to the number of arrival photons, if we take the Q x r distribution as a function of r, the distribution is expected to be linear in the logarithm scale. The fluctuation is caused by the limited statistics, gain variations of PMTs and the depth dependence of ice properties. Because the AMASIM simulates the depth dependence of ice properties and gain variations of PMTs, it is valuable to compare it with MC. The results are shown in Figure 15. Where, the moving threshold method is used and the Npe expectation of MC is obtained from the true Npe in order to see only ice property effect eliminating the gain variation of PMT (Scaled for the comparison). There is good agreement between Data and MC in the low Q region (r > 100~150m). Also the flat distribution due to saturation is observed below r=125~150 m.
As you see in the Figure 14. The saturation in the real Data
is
more serious than MC simulation. The measured Q value for the
brightest
OMs is roughly ~10 times smaller than MC. It may indicate the
weak
point of our simulation for the saturation effect. In other words, our
expectation on the dynamic range in the previous chapter (Figure 9)
should be ~10 times overestimated.
We are trying to develop this restricted dynamic range using the after
pulse information. We can extract the after pulse component in
the
given waveform using the time difference between primary pulse and
secondaries. Since after pulse probability is well known, the original
primary charge information can be estimated from the measured charge of
the after pulse and
its probability (See Figure 15).
Figure 15 : Estimation of original primary charge using after pulse.
We integrated the charge of the secondary pulses which have the time interval 4 to Tmax (micro seconds) after the primary leading edge, where Tmax is the maximum time interval up to the available maximum time bin for given waveform. The Tmax is determined by the time difference between the leading edge of the primary pulse and the last time bin (10240 ns) of TWR. The after pulse probabilities in this time range are obtained for each OMs in the waveform-by-waveform basis. Figure 16 shows the results. We see a big improvement in Q measurement. No saturations is seen using after pulse method. Also there is very good agreement between MC expectation and the after pulse result.
Figure 16 : Measured Q using After Pulse.
However we should note that the Q in Figure 16 is just taken in
average. Because of its small probability, the normal OMs have no after
pulse in
the most of all events. The only very bright OM or very big primary
pulse
can make enough number of after pulses so that we can get the
reasonable
size of error. Figure 17 shows the Q distribution of OM 500 measured by
the after pulse method. The OM 500 is the brightest OM and about 25% of
error is achieved.
Figure 17 : Q distribution of OM 500 measured by the after pulse method
Since the 2002-2003 Summer season, TWR system has been
available for most of all OMs. Figure 18 shows Npe x r
distributions for N2 laser data with all TWR available OMs. Top
figure is for electrical channels and the bottom is optical channels.
The dynamic range with TWR After pulse measurement for electric channel
is about two order of magnitude higher than Muon Daq while the Npe with
TWR primary pulse measurement is limited by a saturation due to a
limited voltage window of TWR digitization. For electrical
channels, the dynamic range with TWR After pulse is extendeds upto more
than 2000 Npes which is about three order of magnitude higher than Muon
Daq while primary pulse for optical channels is about an order of
magnitude higher than Muon Daq.
Base on the prototype TWR system we developed the charge
reconstruction method for AMANDA experiment. Several methods such as
the fixed baseline method, the moving baseline method and after pulse
method, have been tested. For the normal channels and events with the
stable baseline, both results of the fixed and moving baseline method
shows good agreement. But in the
channels with large overshoot, the difference between two method become
large. For the very bright OM with huge Npe, the saturation limitation
of ORB is the most important factor in the Q measurement rather than
which method
is chosen. Even TWR system is helpful to increase the dynamic rage, the
saturation effect in the real data is more serious than MC expectation.
The after pulse method provides an improvement to overcome the
saturation
effect. For very bright OMs no saturation is observed using this
method.
The moving threshold method is accompanied by after pulse method. Since
the baselines is usually distorted by large signal, more accurate
measurement
is available when the moving baseline method is used to correct it.
For the after pulse measurement the counting method should be tried in
the
future. Because the small population of after pulse, the individual
waveforms
are well separated each other. As a homework in the moving threshold
method
the stability and fast process is required. This study should
extend
to the electrical channels after all channels are equipped with TWR
system.